SOP for Calibration of UV Meter

Posted on By

UV Meter Calibration Procedure

Purpose

The purpose of this Standard Operating Procedure (SOP) is to establish the procedures for the calibration of UV meters used in photostability studies and other UV light intensity measurements. Proper calibration ensures accurate and reliable readings.

Scope

This SOP applies to all UV meters used for measuring ultraviolet light intensity within the facility.

Responsibilities

  • Stability Manager: Oversee the calibration process and ensure compliance with this SOP.
  • Calibration Technician: Perform the calibration of UV meters and document the results.
  • Quality Assurance (QA) Team: Review and approve calibration records and reports.

Procedure

  1. Preparation:
    • Ensure that the UV meter is clean and in good working condition.
    • Gather all necessary calibration
equipment, including a certified UV light source with a known intensity.
  • Verify that the calibration equipment is within its calibration due date and properly certified.
  • Familiarize yourself with the UV meter’s user manual and calibration procedures.
  • Initial Check:
    • Turn on the UV meter and allow it to stabilize

      “Ensuring Product Longevity: A Comprehensive Guide to Pharmaceutical Stability Testing for Global Market Success”

      according to the manufacturer’s instructions.
    • Check the battery level and replace if necessary.
    • Ensure that the sensor is clean and free from any obstructions.
  • Calibration Process:
    • Place the UV meter’s sensor at a specified distance from the certified UV light source.
    • Activate the certified UV light source and allow it to reach a stable output.
    • Record the UV light intensity reading displayed by the UV meter.
    • Compare the UV meter reading with the known intensity of the certified UV light source.
    • If the reading is within the acceptable tolerance range, the UV meter is considered calibrated. If not, adjust the UV meter according to the manufacturer’s instructions and repeat the measurement.
  • Adjustment (if necessary):
    • Refer to the UV meter’s user manual for adjustment procedures.
    • Make the necessary adjustments to bring the UV meter readings within the acceptable tolerance range.
    • Repeat the calibration process to verify accuracy after adjustments.
  • Verification and Documentation:
    • Record all calibration data, including:
      • Date of calibration
      • Calibration equipment used
      • UV light intensity readings before and after adjustment (if any)
      • Signatures of the personnel performing and reviewing the calibration
    • Submit the calibration records to the Stability Manager and QA Team for review and approval.
  • Post-Calibration:
    • Turn off the UV meter and store it in a clean, dry place.
    • Ensure that the UV meter is stored in a protective case to avoid damage.
    • Report any deviations or issues encountered during calibration to the Stability Manager and QA Team.
  • Record Keeping:
    • Maintain all calibration records, including reports and certificates of calibration equipment, in a centralized location for easy access and review.
    • Ensure electronic data is backed up regularly and securely stored.
    • Archive records according to the facility’s document retention policy and regulatory requirements.
  • Abbreviations Used

    • QA: Quality Assurance
    • SOP: Standard Operating Procedure
    • UV: Ultraviolet

    Documents

    • Calibration Records
    • Calibration Equipment Certificates
    • Calibration Reports

    References

    • UV Meter User Manual
    • Manufacturer’s Calibration Guidelines

    SOP Version

    Version 1.0 – Effective Date: [Insert Date]

    Related Topics:

    • Outsourced Stability Storage and Testing Procedures:… Outsourced Stability Storage and Testing Procedures: Compliance and Best Practices Outsourced Stability Storage and Testing Procedures: Compliance and Best Practices…
    • Managing Packaging Stability Studies for High-Potency APIs Managing Packaging Stability Studies for High-Potency APIs Managing Packaging Stability Studies for High-Potency APIs Introduction High-potency active pharmaceutical ingredients (HPAPIs)…
    • Stability Testing Requirements: A Comprehensive… Stability Testing Requirements: A Comprehensive Guide for Pharmaceutical Products Stability Testing Requirements: Ensuring Pharmaceutical Product Quality and Compliance Introduction Stability…
    • SOP for Review of Temperature and Humidity Recording Charts Monitoring and Management of Temperature and Humidity Records Purpose The purpose of this Standard Operating Procedure (SOP) is to establish…
    • Stability Studies: Key Regulatory Guidelines for… Pharma Stability Studies: Regulatory Guidelines The pharmaceutical industry operates under stringent quality standards to ensure that every product reaching patients…
    • Stability Testing: A Cornerstone of Pharmaceutical… Overview of Stability Testing in Pharmaceuticals Stability testing is a critical component of pharmaceutical development, ensuring that drugs and medicinal…
    Stability Studies SOP Tags:Accelerated stability, Accelerated stability testing, API degradation products,, Biopharmaceutical stability, Combination product stability,, Container closure integrity testing, Degradation pathways, Drug degradation, Drug degradation pathways, Drug formulation stability, Drug product quality, Drug stability, Drug stability studies,, Forced degradation, Forced degradation studies, GMP and stability studies,, ICH guidelines, ICH guidelines for stability, ICH stability guidelines,, Long-term stability, Long-term stability studies, Long-term stability testing, Parenteral dosage form stability, Pharmaceutical industry, Pharmaceutical products, Pharmaceutical stability, Pharmaceutical stability testing,, Photostability testing, Photostability,, Real-time stability, Real-time stability studies, Regulatory compliance, Regulatory guidelines, Regulatory requirements, Regulatory requirements stability studies, Regulatory submissions, Shelf life determination, Shelf life extension, Shelf life,, Short-term stability, Stability analysis, Stability chamber conditions, Stability chambers, Stability data analysis, Stability data interpretation, Stability indicating assays Stability indicating methods, Stability indicating methods, Stability protocol, Stability protocols, Stability studies, Stability study acceptance criteria,, Stability study best practices, Stability study data analysis, Stability study data trending,, Stability study design, Stability study deviations, Stability study documentation,, Stability study guidelines, Stability study photostability,, Stability study procedure,, Stability study protocol, Stability study protocol design, Stability study report,, Stability study sample handling, Stability study SOPs, Stability testing, Stability testing best practices, Stability testing challenges, Stability testing compliance, Stability testing conditions, Stability testing for biologics, Stability testing for drug approval, Stability testing for generic drugs, Stability testing for new drug products,, Stability testing for pharmaceuticals,, Stability testing guidelines, Stability testing innovations, Stability testing methods, Stability testing of APIs,, Stability testing of finished products,, Stability testing parameters, Stability testing protocols,, Stability testing regulations, Stability testing requirements, Stability testing trends, Stability testing validation, Statistical analysis stability, Storage conditions for stability testing,, Temperature and humidity control

    Post navigation

    Previous Post: SOP for Freeze Thaw Studies
    Next Post: SOP for Operation of UV Meter

    Quick Guide

    • Stability Tutorials
    • Stability Testing Types
      • Types of Stability Studies
      • Real-Time and Accelerated Stability Studies
      • Intermediate and Long-Term Stability Testing
      • Freeze-Thaw and Thermal Cycling Studies
      • Photostability and Oxidative Stability Studies
      • Stability Testing for Biopharmaceuticals
    • Stability Studies SOP
    • ‘How to’ – Stability Studies
    • Regulatory Guidelines
    • Shelf Life and Expiry Dating
    • Stability Documentation
    • Stability Studies – API
    • Stability Studies Blog
    • Stability Studies FAQ
    • Packaging – Containers – Closers
    Widget Image
    • Maintain Backup Stability Chambers to Prevent Data Loss in Case of Failure

      Understanding the Tip: Why backup chambers are essential: Stability chambers are critical infrastructure in pharmaceutical QA.
      A sudden malfunction—due to power failure, temperature controller breakdown,… Read more

    Copyright © 2025 StabilityStudies.in.

    Powered by PressBook WordPress theme